Справка по командам SCPI
640 FTB-5240S/S-P/BP
Product-Specific Commands—Description
:UNIT[1..n]:RATio?
Description This query returns the selected default unit for
commands which program relative power.
At *RST, default relative power unit is set to DB.
Syntax :UNIT[1..n]:RATio?
Parameter(s) None
Response Syntax <Unit>
Response(s) Unit:
The response data syntax for <Unit> is defined as
a <CHARACTER RESPONSE DATA> element.
The response corresponds to the selected default
relative power unit.
DB: dB power unit is selected.
W/W: watt ratio power unit is selected.
%: percent power unit is selected.
Example(s) UNIT:RAT W/W
UNIT:RAT? Returns W/W
See Also :UNIT[1..n]:POWer
:UNIT[1..n]:SPECtrum
:UNIT[1..n]:RATio