Справка по командам SCPI
OSA 601
Product-Specific Commands—Description
:SENSe[1..n]:FREQuency:STOP
Description This command sets instrument sweep stop
frequency.
At *RST, this value is set to 196.5852 THz.
Syntax :SENSe[1..n]:FREQuency:STOP<wsp><Stop[<w
sp>HZ]>|MAXimum|MINimum|DEFault
Parameter(s) Stop:
The program data syntax for <Stop> is defined as
a <numeric_value> element followed by an
optional <SUFFIX PROGRAM DATA> element.
The allowed <SUFFIX PROGRAM DATA>
element is HZ. The <Stop> special forms
MINimum, MAXimum and DEFault are accepted
on input.
MINimum allows to set the instrument to the
smallest supported value.
MAXimum allows to set the instrument to the
greatest supported value.
DEFault allows the instrument to select a value
for the <Stop> parameter.
The <Stop> parameter corresponds to a valid
sweep stop frequency value.
The SENSe[1..n]:FREQuency:STOP? MIN and
SENSe[1..n]:FREQuency:STOP? MAX queries
can be used to determine valid sweep stop
frequency range.