Exfo FTB -5240S/5240BP Optical Spectrum Analyzer for FT Manual de usuario Pagina 645

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 688
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 644
Справка по командам SCPI
OSA 639
Product-Specific Commands—Description
:UNIT[1..n]:RATio
Description This command selects a default unit for
commands which program relative power.
At *RST, default relative power unit is set to DB.
Syntax :UNIT[1..n]:RATio<wsp>DB|W/W|PCT
Parameter(s) Unit:
The program data syntax for the first parameter is
defined as a <CHARACTER PROGRAM DATA>
element. The allowed <CHARACTER PROGRAM
DATA> elements for this parameter are:
DB|W/W|PCT.
The parameter corresponds to the newly selected
default relative power unit.
DB: selects dB power unit.
W/W: selects watt ratio power unit.
PCT: selects percent power unit
Example(s) UNIT:POW W
UNIT:POW? Returns W
UNIT:RAT DB
UNIT:RAT? Returns DB
UNIT:POW? Returns DBM
Notes Changing default relative power unit also sets the
default absolute power unit (UNIT:POWer) to the
corresponding setting.
See Also :UNIT[1..n]:POWer
:UNIT[1..n]:SPECtrum
:UNIT[1..n]:RATio?
Vista de pagina 644
1 2 ... 640 641 642 643 644 645 646 647 648 649 650 ... 687 688

Comentarios a estos manuales

Sin comentarios