Справка по командам SCPI
OSA 405
Product-Specific Commands—Description
:CALCulate[1..n][:WDM]:CHANnel:AUTO:
NOISe:AUTO
Description This command controls activation of i-InBand
noise measurement for WDM analysis default
channel.
At *RST, auto noise is set to off (disabled).
Syntax :CALCulate[1..n][:WDM]:CHANnel:AUTO:NOISe:
AUTO<wsp><Auto>
Parameter(s) Auto:
The program data syntax for <Auto> is defined as
a <Boolean Program Data> element. The <Auto>
special forms ON and OFF are accepted on input
for increased readability. ON corresponds to 1
and OFF corresponds to 0.
The <Auto> parameter corresponds to the new
state of auto noise measurement.
0 or OFF: disable default channel auto noise.
1 or ON: enable default channel auto noise.