Exfo FTB -5240S/5240BP Optical Spectrum Analyzer for FT Manual de usuario Pagina 418

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 688
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 417
Справка по командам SCPI
412 FTB-5240S/S-P/BP
Product-Specific Commands—Description
:CALCulate[1..n][:WDM]:CHANnel:AUTO:
NOISe:DISTance:[WAVelength]
Description This command sets the wavelength distance from
peak to center of noise region for noise
measurement of the WDM analysis default
channel.
At *RST, default channel noise measurement
distance is set to 100.0 GHz.
Syntax :CALCulate[1..n][:WDM]:CHANnel:AUTO:NOISe:
DISTance:[WAVelength]<wsp><Distance[<wsp>
M]>|MAXimum|MINimum|DEFault
Parameter(s) Distance:
The program data syntax for <Distance> is
defined as a <numeric_value> element followed
by an optional <SUFFIX PROGRAM DATA>
element. The allowed <SUFFIX PROGRAM
DATA> element is M. The <Distance> special
forms MINimum, MAXimum and DEFault are
accepted on input.
MINimum allows to set the instrument to the
smallest supported value.
MAXimum allows to set the instrument to the
greatest supported value.
Vista de pagina 417
1 2 ... 413 414 415 416 417 418 419 420 421 422 423 ... 687 688

Comentarios a estos manuales

Sin comentarios