Exfo FTB-500 (Window XP) Manual de usuario Pagina 272

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 318
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 271
Справка по командам IEEE 488.2 и специальным командам
260 FTB-500
Специальные команды
:INSTrument:CATalog:FULL?
Description The INSTrument:CATalog:FULL? returns a list of
<STRING RESPONSE DATA> - <NR1 NUMERIC
RESPONSE DATA> pairs. The <STRING
RESPONSE DATA> contains the name of the
logical instrument. The immediately following
<NR1 NUMERIC RESPONSE DATA> formatted
number is its associated logical instrument
number. All response data elements are
separated by commas. If no logical instrument is
defined, a null <STRING RESPONSE DATA>
value followed by a zero is returned.
This is not affected by a *RST command.
Syntax :INSTrument:CATalog:FULL?
Parameter(s) None
Response Syntax <Catalog>
Response(s) Catalog:
The response data syntax for <Catalog> is
defined as a <STRING RESPONSE DATA>
element.
The list of <STRING RESPONSE DATA>
contains the names of all logical instruments and
groups. The immediately following <NR1
NUMERIC RESPONSE DATA> formatted number
is its associated logical instrument number.
Example(s) INST:CAT:FULL? returns “IQS-2403 WDM Laser
Source”,10,”IQS-2403 WDM Laser Source”,11
See Also INSTrument:CATalog?.
Vista de pagina 271
1 2 ... 267 268 269 270 271 272 273 274 275 276 277 ... 317 318