PM-200 BUILT-IN POWER METER SPECIFICATIONS
d
Calibrated wavelengths (nm) 850, 1300, 1310, 1490, 1550, 1625, 1650
Optional CWDM calibrated
wavelengths (nm)
1270, 1290, 1310, 1330, 1350, 1370, 1390, 1410, 1430, 1450, 1470, 1490,
1510, 1530, 1550, 1570, 1590, 1610, 1383, 1625
Power range (dBm) 10 to —86 (InGaAs)
26 to —64 (GeX)
Uncertainty (%)
e
±5 % ± 3 pW (InGaAs)
±5 % ± 0.4 nW (GeX)
Display resolution (dB)
InGaAs
GeX
0.01 = max to —76 dBm
0.1 = —76 dBm to —86 dBm
1 = —86 dBm to min
0.01 = max to —54 dBm
0.1 = —54 dBm to —64 dBm
1 = —64 dBm to min
Automatic offset nulling range
f
Max power to —63 dBm for InGaAs
Max power to —40 dBm for GeX
Tone detection (Hz) 270/1000/2000
Notes
a. All specifications valid at 23
o
C (73
o
F).
b. Not including internal batteries. Battery maximum storage temperature: 60
o
C (140
o
F).
c. The selected model may change without notice.
d. At 23 ºC ± 1 ºC, 1550 nm and FC connector. With modules in idle mode. Battery operated.
e. Up to 5 dBm.
f. For ±0.05 dB, from 18 ºC to 28 ºC.
LASER RADIATION
AVOID DIRECT EYE EXPOSURE
CLASS 3R LASER PRODUCT
IEC 60825-1:2007
21 CFR 1040.10
l: 650 ±10 nm
P
out
maximum < 5mW (into free space)
Comentarios a estos manuales