Remote Control
126 FLS-2600B
Other Commands
Other Commands
SOURce:SWEep:SAVE
Description This function is used to save the following sweep setup
parameters:
start wavelength, end wavelength, mode, speed, length,
cycles, reverse, trigger for continuous sweep
start wavelength, end wavelength, mode, step, pause,
cycles, reverse, trigger for step-by-step sweepYou can save
up to five configurations.
Syntax SOUR:SWE:SAVE<wsp><number>
Parameters The <number> parameter represents the number you want to
use in the Save list. You can choose a number from 1 to 5.
Example SOUR:SWE:SAVE 2
Note You cannot use this command during a sweep.
SOURce:SWEep:RECAll
Description This function is used to recall a sweep setup saved in your list.
Syntax SOUR:SWE:RECA<wsp><number>
Parameters The <number> parameter represents the number of the item
you want to recall. You can choose a number from 1 to 5.
Example SOUR:SWE:RECA 3
Note You cannot use this command during a sweep.