Exfo BFA-3000 Bare Fiber Adapter Manual de usuario

Busca en linea o descarga Manual de usuario para Instrumentos de medición Exfo BFA-3000 Bare Fiber Adapter. EXFO BFA-3000 Bare Fiber Adapter User Manual Manual de usuario

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 2
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 0
BFA-3000 Universal Bare Fiber Adapter
Instruction Sheet
Bare fiber adapters (BFA) connect single-core, single-mode or multimode bare fibers to optical
power meter detector ports in a secure manner.
To connect a fiber with the BFA-3000:
1. Remove 80 mm (3.15 in) of outer cable jacket and kevlar material.
2. Strip 30 mm (1.18 in) of buffer, then clean the fiber with alcohol or acetone.
3. Cleave the fiber. Leaving 18 mm (0.2 in) of bare fiber will ensure an optimal fit in the
BFA-3000.
4. Squeeze open the BFA-3000 by applying pressure with your thumb and index finger.
5. Place the fiber inside the opening, leaving a small length of fiber out. The fiber must be
inserted into the V groove at the front of the BFA-3000.
80 mm (3,15 in)
18 mm
(0.2 in)
Fiber
V groove
Vista de pagina 0
1 2

Indice de contenidos

Pagina 1 - Instruction Sheet

BFA-3000 Universal Bare Fiber Adapter Instruction SheetBare fiber adapters (BFA) connect single-core, single-mode or multimode bare fibers to optical

Pagina 2

© 2011 EXFO Inc. All rights reserved. Printed in Canada. P/N: 1060915Technical Support Group400 Godin AvenueQuébec (Quebec) G1M 2K2CANADA1 866 683-015

Comentarios a estos manuales

Sin comentarios