Exfo FLS-2600B Tunable Laser Source Manual de usuario Pagina 102

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 157
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 101
Remote Control
94 FLS-2600B
Specific Commands
OUTPut:STATe?
Description This query returns the current source state.
Syntax OUTP:STAT?
Response A boolean value representing the current source state:
0: the source is deactivated.
1: the source is activated.
See also OUTP:STAT, OUTP:MODE, OUTP:MODE? and OUTP:ASE?
SOURce:MODE
Description This command is used to toggle between the normal and
high-resolution modes.
Syntax SOUR:MODE<wsp><mode>
Parameters The <mode> parameter is a boolean value representing the
mode state:
0: Normal mode
1: High-Resolution mode
Example SOUR:MODE 1
Note This command cannot be processed if your tunable laser
source is performing a sweep.
See also SOUR:MODE?
Vista de pagina 101
1 2 ... 97 98 99 100 101 102 103 104 105 106 107 ... 156 157

Comentarios a estos manuales

Sin comentarios