Remote Control
92 FLS-2600B
Specific Commands
OUTPut:ASE?
Description This query returns a value indicating whether the ASE option is
available on the tunable laser source.
Syntax OUTP:ASE?
Response A boolean value indicating whether the ASE option is available
on the FLS-2600B Tunable Laser Source:
0: the ASE option is not available
1: the ASE option is available
Note The response will always be “not available”.
See also OUTP:MODE, OUTP:MODE?, OUTP:STAT and OUTP:STAT?
OUTPut:MODE
Description This command is used to set the source mode (tunable or ASE).
Syntax OUTP:MODE<wsp><mode>
Parameters The <mode> parameter is a boolean parameter representing
the source mode:
0: Tunable
1: ASE
Note This command cannot be performed if the FLS-2600B is
performing a sweep. It will return an error if ASE is selected
because it is not available.
Example OUTP:MODE 0
See also OUTP:ASE?, OUTP:MODE?, OUTP:STAT and OUTP:STAT?