SCPI Command Reference
Variable Attenuator 91
Product-Specific Commands—Description
:CONTrol[1..n]:MODE
Description This command selects the attenuator's control
mode.
At *RST, the control mode is ATTenuation.
Syntax :CONTrol[1..n]:MODE<wsp>ATTenuation|
POWer
Parameter(s) Parameter 1:
The program data syntax for the first parameter is
defined as a <CHARACTER PROGRAM DATA>
element. The allowed <CHARACTER PROGRAM
DATA> elements for this parameter are:
ATTenuation|POWer.
This parameter represents the newly selected
control mode.
ATTenuation: selects the attenuation control
mode.
POWer:selects the power control mode.
Example(s) CONT:MODE POW
CONT:MODE ATTENUATION
See Also :CONTrol[1..n]:MODE?
:INPut[1..n]:ATTenuation
:INPut[1..n]:RATTenuation
:OUTPut[1..n]:POWer
:OUTPut[1..n]:RPOWer