Exfo IQS-3150 Variable Attenuator for IQS-500/600 Manual de usuario Pagina 145

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 161
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 144
SCPI Command Reference
Variable Attenuator 139
Product-Specific Commands—Description
:OUTPut[1..n]:RPOWer?
Description This query returns a value indicating either the
current or the minimum/maximum relative
power value.
At *RST, the value that will be set depends on the
instrument you have.
Syntax :OUTPut[1..n]:RPOWer?[<wsp>MAXimum|MINi
mum|DEFault]
Parameter(s) Parameter 1:
The program data syntax for the first parameter is
defined as a <CHARACTER PROGRAM DATA>
element. The allowed <CHARACTER PROGRAM
DATA> elements for this parameter are:
MAXimum|MINimum|DEFault.
MINimum is used to retrieve the instrument's
lowest supported value.
MAXimum is used to retrieve the instrument's
highest supported value.
DEFault is used to retrieve the instrument's
default value.
Response Syntax <RelativePower>
Response(s) RelativePower:
The response data syntax for <RelativePower>
is defined as a <NR3 NUMERIC RESPONSE
DATA> element.
The <RelativePower> parameter corresponds to
a valid relative power value.
Vista de pagina 144
1 2 ... 140 141 142 143 144 145 146 147 148 149 150 ... 160 161

Comentarios a estos manuales

Sin comentarios