Exfo IQS-3150 Variable Attenuator for IQS-500/600 Manual de usuario Pagina 150

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 161
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 149
SCPI Command Reference
144 IQS-3150
Product-Specific Commands—Description
:SENSe[1..n]:CORRection:COLLect:
ZERO
Description This command performs an offset nulling on the
internal power meter.
This command is an event and has no associated
*RST condition or query form.
Syntax :SENSe[1..n]:CORRection:COLLect:ZERO
Parameter(s) None
Example(s) STAT? Must return READY
SENS:CORR:COLL:ZERO
STAT:OPER:BIT10:COND? Keep resending the
query as long as the operation is not complete
(returned value is not 0).
Notes This command will take at least 3 seconds to
complete.
See Also :CALibration[1..n]:ZERO
:STATus:OPERation:BIT<n>:CONDition?
Vista de pagina 149
1 2 ... 145 146 147 148 149 150 151 152 153 154 155 ... 160 161

Comentarios a estos manuales

Sin comentarios