SCPI Command Reference
124 IQS-3150
Product-Specific Commands—Description
:OUTPut[1..n]:LOCK[:STATe]?
Description This query returns the lock state of the
instrument's shutter. If the shutter is closed using
the red push button (located on the front panel of
the instrument), the shutter is automatically
locked. Once the shutter is locked, it can only be
unlocked by using the red push button again.
At *RST the lock state remains unchanged.
Syntax :OUTPut[1..n]:LOCK[:STATe]?
Parameter(s) None
Response Syntax <ShutterLockState>
Response(s) ShutterLockState:
The response data syntax for
<ShutterLockState> is defined as a <NR1
NUMERIC RESPONSE DATA> element.
The <ShutterLockState> corresponds to the
current lock state of the shutter.
0: The shutter is unlocked and can be controlled
using the OUTup[1..n][:STATe] command.
1: The shutter is locked and can not be controlled
remotely. In this case, the shutter can only be
unlocked by using the red push button (located
on the instrument’s front panel).
Example(s) OUTP:LOCK:STAT? Returns: 0 (The shutter is
unlocked and can be controlled using the
OUTup[1..n][:STATe] command.)
See Also :OUTPut[1..n][:STATe]