User GuideVariable Attenuator for IQS PlatformsIQS-3150

Pagina 2

Introducing the IQS-3150 Variable Attenuator4 IQS-3150Main Software FeaturesMain Software FeaturesThe dedicated software functions are very similar wh

Pagina 3 - Contents

SCPI Command Reference94 IQS-3150Product-Specific Commands—Description:INPut[1..n]:ATTenuationDescription This command sets the absolute attenuation t

Pagina 4

SCPI Command ReferenceVariable Attenuator 95Product-Specific Commands—DescriptionDEFault allows the instrument to select a value for the <Attenuati

Pagina 5 - Certification Information

SCPI Command Reference96 IQS-3150Product-Specific Commands—Description:INPut[1..n]:ATTenuation?Description This query returns a value indicating eithe

Pagina 6 - DECLARATION OF CONFORMITY

SCPI Command ReferenceVariable Attenuator 97Product-Specific Commands—DescriptionResponse(s) Attenuation: The response data syntax for <Attenuation

Pagina 7 - 1 Introducing the IQS-3150

SCPI Command Reference98 IQS-3150Product-Specific Commands—Description:INPut[1..n]:OFFSetDescription This command sets an offset value for the attenua

Pagina 8 - Local and Remote Control

SCPI Command ReferenceVariable Attenuator 99Product-Specific Commands—DescriptionExample(s) INP:WAV 1310 NMCONT:MODE ATTOUTP:APM ABSINP:OFFS DEFINP:AT

Pagina 9 - Typical Applications

SCPI Command Reference100 IQS-3150Product-Specific Commands—Description:INPut[1..n]:OFFSet?Description This query returns a value indicating either th

Pagina 10 - Main Software Features

SCPI Command ReferenceVariable Attenuator 101Product-Specific Commands—DescriptionResponse(s) Offset: The response data syntax for <Offset> is d

Pagina 11 - Control Mode Description

SCPI Command Reference102 IQS-3150Product-Specific Commands—Description:INPut[1..n]:RATTenuationDescription This command sets the relative attenuation

Pagina 12 - Display Modes

SCPI Command ReferenceVariable Attenuator 103Product-Specific Commands—DescriptionDEFault allows the instrument to select a value for the <Relative

Pagina 13 - Multiple Module Control

Introducing the IQS-3150 Variable AttenuatorVariable Attenuator 5Main Software FeaturesControl Modes (Self-Adjusting Modules)With the self-adjusting m

Pagina 14 - MPORTANT

SCPI Command Reference104 IQS-3150Product-Specific Commands—DescriptionSTAT:OPER:BIT8:COND? Keep sending the query as long as attenuation is not reach

Pagina 15 - Laser Safety Information

SCPI Command ReferenceVariable Attenuator 105Product-Specific Commands—Description:INPut[1..n]:RATTenuation?Description This query returns either the

Pagina 16

SCPI Command Reference106 IQS-3150Product-Specific Commands—DescriptionThe <RelativeAttenuation> response represents either the current or the M

Pagina 17 - Variable Attenuator

SCPI Command ReferenceVariable Attenuator 107Product-Specific Commands—Description:INPut[1..n]:REFerenceDescription This command sets, for the current

Pagina 18 - 12 IQS-3150

SCPI Command Reference108 IQS-3150Product-Specific Commands—DescriptionExample(s) INP:WAV 1310 NMCONT:MODE ATTOUTP:APM ABSINP:OFFS 0.000 DBINP:RATT 33

Pagina 19 - Retaining screw knob

SCPI Command ReferenceVariable Attenuator 109Product-Specific Commands—Description:INPut[1..n]:REFerence?Description This query returns either the cur

Pagina 20 - 14 IQS-3150

SCPI Command Reference110 IQS-3150Product-Specific Commands—DescriptionResponse(s) Reference: The response data syntax for <Reference> is define

Pagina 21 - Variable Attenuator 15

SCPI Command ReferenceVariable Attenuator 111Product-Specific Commands—Description:INPut[1..n]:WAVelengthDescription This command selects a specific w

Pagina 22 - 16 IQS-3150

SCPI Command Reference112 IQS-3150Product-Specific Commands—DescriptionDEFault allows the instrument to select a value for the <Wavelength> para

Pagina 23 - Data Display

SCPI Command ReferenceVariable Attenuator 113Product-Specific Commands—Description:INPut[1..n]:WAVelength?Description This query returns a value indic

Pagina 24 - Status Bar

Introducing the IQS-3150 Variable Attenuator6 IQS-3150Main Software FeaturesDisplay ModesBoth standard and self-adjusting modules give you a choice of

Pagina 25

SCPI Command Reference114 IQS-3150Product-Specific Commands—DescriptionResponse(s) Wavelength: The response data syntax for <Wavelength> is defi

Pagina 26 - Exiting the Application

SCPI Command ReferenceVariable Attenuator 115Product-Specific Commands—Description:OUTPut[1..n]:ALC[:STATe]Description This command activates or deact

Pagina 27 - Attenuator

SCPI Command Reference116 IQS-3150Product-Specific Commands—Description:OUTPut[1..n]:ALC[:STATe]?Description This query indicates if the power trackin

Pagina 28 - Basic Settings

SCPI Command ReferenceVariable Attenuator 117Product-Specific Commands—Description:OUTPut[1..n]:APModeDescription This command selects, for the active

Pagina 29 - Setting the B Value

SCPI Command Reference118 IQS-3150Product-Specific Commands—DescriptionExample(s) INP:WAV 1310 NMCONT:MODE ATTOUTP:APM ABSINP:RATT 42.75INP:RATT? Ret

Pagina 30 - 24 IQS-3150

SCPI Command ReferenceVariable Attenuator 119Product-Specific Commands—Description:OUTPut[1..n]:APMode?Description This query returns, for active cont

Pagina 31 - Variable Attenuator 25

SCPI Command Reference120 IQS-3150Product-Specific Commands—Description:OUTPut[1..n]:DTOleranceDescription This command specifies the drift tolerance

Pagina 32 - 26 IQS-3150

SCPI Command ReferenceVariable Attenuator 121Product-Specific Commands—DescriptionDEFault allows the instrument to select a value for the <Drift>

Pagina 33 - Setting the Drift Tolerance

SCPI Command Reference122 IQS-3150Product-Specific Commands—Description:OUTPut[1..n]:DTOlerance?Description This query returns the drift tolerance tha

Pagina 34 - 28 IQS-3150

SCPI Command ReferenceVariable Attenuator 123Product-Specific Commands—DescriptionResponse(s) Drift: The response data syntax for <Drift> is def

Pagina 35 - Variable Attenuator 29

Introducing the IQS-3150 Variable AttenuatorVariable Attenuator 7Main Software FeaturesOutput powerMultiple Module ControlThe multimodule application

Pagina 36

SCPI Command Reference124 IQS-3150Product-Specific Commands—Description:OUTPut[1..n]:LOCK[:STATe]?Description This query returns the lock state of the

Pagina 37 - The selected

SCPI Command ReferenceVariable Attenuator 125Product-Specific Commands—Description:OUTPut[1..n]:OFFSetDescription This command sets a power offset val

Pagina 38 - 32 IQS-3150

SCPI Command Reference126 IQS-3150Product-Specific Commands—DescriptionDEFault allows the instrument to select a value for the <Offset> paramete

Pagina 39 - Variable Attenuator 33

SCPI Command ReferenceVariable Attenuator 127Product-Specific Commands—Description:OUTPut[1..n]:OFFSet?Description This query returns a value indicati

Pagina 40 - Defining an Offset Value

SCPI Command Reference128 IQS-3150Product-Specific Commands—DescriptionResponse(s) Offset: The response data syntax for <Offset> is defined as a

Pagina 41

SCPI Command ReferenceVariable Attenuator 129Product-Specific Commands—Description:OUTPut[1..n]:POWerDescription This command sets the absolute output

Pagina 42 - Creating a Test Sequence

SCPI Command Reference130 IQS-3150Product-Specific Commands—DescriptionExample(s) INP:WAV 1310 NMCONT:MODE POWOUTP:POW -15.000 DBMSTAT:OPER:BIT8:COND

Pagina 43

SCPI Command ReferenceVariable Attenuator 131Product-Specific Commands—Description:OUTPut[1..n]:POWer?Description This query returns a value indicatin

Pagina 44 - Setting a Starting Time

SCPI Command Reference132 IQS-3150Product-Specific Commands—DescriptionResponse(s) Power: The response data syntax for <Power> is defined as a &

Pagina 45

SCPI Command ReferenceVariable Attenuator 133Product-Specific Commands—Description:OUTPut[1..n]:REFerenceDescription This command sets a power referen

Pagina 46 - Setting Repetition Scheme

Introducing the IQS-3150 Variable Attenuator8 IQS-3150ConventionsConventionsBefore using the product described in this manual, you should understand t

Pagina 47

SCPI Command Reference134 IQS-3150Product-Specific Commands—DescriptionDEFault allows the instrument to select a value for the <Reference> param

Pagina 48 - Managing Sequence Files

SCPI Command ReferenceVariable Attenuator 135Product-Specific Commands—Description:OUTPut[1..n]:REFerence?Description This query returns either the cu

Pagina 49

SCPI Command Reference136 IQS-3150Product-Specific Commands—DescriptionResponse(s) Reference: The response data syntax for <Reference> is define

Pagina 50 - 44 IQS-3150

SCPI Command ReferenceVariable Attenuator 137Product-Specific Commands—Description:OUTPut[1..n]:RPOWerDescription This command sets the relative power

Pagina 51

SCPI Command Reference138 IQS-3150Product-Specific Commands—Descriptiona) In ABSolute mode, <RelativePower> = absolute power + power offset valu

Pagina 52

SCPI Command ReferenceVariable Attenuator 139Product-Specific Commands—Description:OUTPut[1..n]:RPOWer?Description This query returns a value indicati

Pagina 53

SCPI Command Reference140 IQS-3150Product-Specific Commands—Descriptiona) In ABSOLUTE mode, <RelativePower> = absolute power + power offset valu

Pagina 54 - Nulling Electrical Offsets

SCPI Command ReferenceVariable Attenuator 141Product-Specific Commands—Description:OUTPut[1..n][:STATe]Description This command controls the state of

Pagina 55

SCPI Command Reference142 IQS-3150Product-Specific Commands—Description:OUTPut[1..n][:STATe]?Description This query returns the state of the instrumen

Pagina 56 - 50 IQS-3150

SCPI Command ReferenceVariable Attenuator 143Product-Specific Commands—Description:READ[1..n][:SCALar]:POWer:DC?Description This query returns the pow

Pagina 57 - Variable Attenuator 51

Introducing the IQS-3150 Variable AttenuatorVariable Attenuator 9Laser Safety InformationLaser Safety InformationYour IQS-3150 Variable Attenuator doe

Pagina 58 - (Self-Adjusting Modules)

SCPI Command Reference144 IQS-3150Product-Specific Commands—Description:SENSe[1..n]:CORRection:COLLect:ZERODescription This command performs an offset

Pagina 59

SCPI Command ReferenceVariable Attenuator 145Product-Specific Commands—Description:SNUMber?Description This query returns a value indicating the instr

Pagina 60 - High-Power Detection

SCPI Command Reference146 IQS-3150Product-Specific Commands—Description:STATus?Description This query returns a value indicating the status of the att

Pagina 61 - Attenuators

SCPI Command ReferenceVariable Attenuator 147Product-Specific Commands—Description:STATus:OPERation:BIT<n>:CONDition?Description This query retu

Pagina 62 - Selecting Modules to Control

SCPI Command Reference148 IQS-3150Product-Specific Commands—DescriptionBit <9>: When the returned value is 1, the instrument’s mechanism is bein

Pagina 63

SCPI Command ReferenceVariable Attenuator 149Product-Specific Commands—Description:STATus:QUEStionable:BIT<n>:CONDition?Description This query r

Pagina 64 - 58 IQS-3150

SCPI Command Reference150 IQS-3150Product-Specific Commands—DescriptionBit <10>: When the returned value is 1, the operation temperature is outs

Pagina 65 - Attenuator Modules

IndexVariable Attenuator 151IndexAAbout function tab ... 79Absolutedisplay mode ...

Pagina 66 - Using Monitor Windows

Index152 IQS-3150display modeAbsolute ... 6Reference... 6

Pagina 67

IndexVariable Attenuator 153modulefront panel description ... 1insertion ... 11m

Pagina 69

Index154 IQS-3150saving a sequence file ... 43saving configuration ... 44sequenceadding

Pagina 70

www.EXFO.com · [email protected]/N: 1052363CORPORATE HEADQUARTERS 400 Godin Avenue Quebec (Quebec) G1M 2K2 CANADATel.: 1 418 683-0211 · Fax: 1 418 683-

Pagina 71 - Insertion Loss

Variable Attenuator 112 Getting Started with Your Variable AttenuatorInserting and Removing Test ModulesTo insert a module into the controller or expa

Pagina 72 - 66 IQS-3150

Getting Started with Your Variable Attenuator12 IQS-3150Inserting and Removing Test Modules4. Insert the protruding edges of the module into the groov

Pagina 73 - 8 Maintenance

Getting Started with Your Variable AttenuatorVariable Attenuator 13Inserting and Removing Test ModulesThe module is correctly inserted when its front

Pagina 74

ii IQS-3150Copyright © 2006–2008 EXFO Electro-Optical Engineering Inc. All rights reserved. No part of this publication may be reproduced, stored in a

Pagina 75

Getting Started with Your Variable Attenuator14 IQS-3150Inserting and Removing Test Modules3. Cover empty slots with the supplied protective covers.3a

Pagina 76

Getting Started with Your Variable AttenuatorVariable Attenuator 15Starting the Variable Attenuator ApplicationStarting the Variable Attenuator Applic

Pagina 77

Getting Started with Your Variable Attenuator16 IQS-3150Starting the Variable Attenuator Application2. Click Start Application.ORPress the green LED p

Pagina 78

Getting Started with Your Variable AttenuatorVariable Attenuator 17Starting the Variable Attenuator ApplicationData DisplaySome elements of the data d

Pagina 79 - Calibration)

Getting Started with Your Variable Attenuator18 IQS-3150Starting the Variable Attenuator ApplicationTitle BarThe title bar is located at the top of th

Pagina 80 - Recalibrating the Unit

Getting Started with Your Variable AttenuatorVariable Attenuator 19Entering Values Using Sliders and Numeric BoxesEntering Values Using Sliders and Nu

Pagina 81 - Recycling and Disposal

Getting Started with Your Variable Attenuator20 IQS-3150Exiting the ApplicationExiting the ApplicationClosing any application that is not currently be

Pagina 82

Variable Attenuator 213 Setting Up Your Variable AttenuatorThe IQS-3150 comes with factory default settings; however, you can create custom settings a

Pagina 83 - 9 Troubleshooting

Setting Up Your Variable Attenuator22 IQS-3150Basic SettingsCustomizing the Wavelength and Fine-Tuning Step ListsYou can define lists of wavelengths a

Pagina 84 - IQS-3150-X XX

Setting Up Your Variable AttenuatorVariable Attenuator 23Basic SettingsSetting the B ValueThe B value settings are necessary for the X+B display mode.

Pagina 85 - Variable Attenuator 79

ContentsVariable Attenuator iiiContentsCertification Information ...

Pagina 86 - Transportation

Setting Up Your Variable Attenuator24 IQS-3150Basic SettingsTo set a correction factor for a wavelength:1. On the Settings function tab, click the Pre

Pagina 87 - 10 Warranty

Setting Up Your Variable AttenuatorVariable Attenuator 25Basic Settings3. According to the control mode used, select the correction factor as follows:

Pagina 88 - Certification

Setting Up Your Variable Attenuator26 IQS-3150Basic SettingsInput PowerThe Input Power setting, only available in the Attenuation control mode, allows

Pagina 89 - Service and Repairs

Setting Up Your Variable AttenuatorVariable Attenuator 27Basic SettingsSetting the Drift ToleranceThe drift tolerance setting is only available for th

Pagina 90 - 84 IQS-3150

Setting Up Your Variable Attenuator28 IQS-3150Basic SettingsTo set the drift tolerance:1. On the Settings function tab, click the Presets tab.2. Under

Pagina 91 - A Technical Specifications

Setting Up Your Variable AttenuatorVariable Attenuator 29Control Center Settings and ControlsControl Center Settings and ControlsOnce the basic settin

Pagina 92 - Technical Specifications

Setting Up Your Variable Attenuator30 IQS-3150Control Center Settings and ControlsSelecting a Display ModeThe active display mode determines the signi

Pagina 93

Setting Up Your Variable AttenuatorVariable Attenuator 31Control Center Settings and ControlsTo select a display mode:1. Click the Instrument function

Pagina 94 - Quick Reference Command Tree

Setting Up Your Variable Attenuator32 IQS-3150Control Center Settings and ControlsSelecting a Wavelength and a Fine-Tuning StepThe attenuator waveleng

Pagina 95 - SCPI Command Reference

Setting Up Your Variable AttenuatorVariable Attenuator 33Control Center Settings and ControlsChanging the Attenuation or the Output PowerYou can chang

Pagina 96 - :CALibration[1..n]:ZERO

Contentsiv IQS-31508 Maintenance ...67Cleaning Fixed Conn

Pagina 97 - :CONTrol[1..n]:MODE

Setting Up Your Variable Attenuator34 IQS-3150Control Center Settings and ControlsDefining an Offset ValueThe offset value is not applied to a particu

Pagina 98 - :CONTrol[1..n]:MODE?

Setting Up Your Variable AttenuatorVariable Attenuator 35Creating a Test SequenceCreating a Test SequenceThe IQS-3150 allows you to create test sequen

Pagina 99 - :INPut[1..n]:ARESolution?

Setting Up Your Variable Attenuator36 IQS-3150Creating a Test SequenceAdding, Inserting, and Removing StepsYou can customize test sequences:³ by addin

Pagina 100 - :INPut[1..n]:ATTenuation

Setting Up Your Variable AttenuatorVariable Attenuator 37Creating a Test Sequence3. In the Attenuation or the Output Power box (depending on the selec

Pagina 101

Setting Up Your Variable Attenuator38 IQS-3150Creating a Test SequenceSetting a Starting TimeThe start of a test sequence can be set to a definite tim

Pagina 102 - :INPut[1..n]:ATTenuation?

Setting Up Your Variable AttenuatorVariable Attenuator 39Creating a Test Sequence3. Under Starting Time, click the Relative or the Absolute button.4.

Pagina 103

Setting Up Your Variable Attenuator40 IQS-3150Creating a Test SequenceSetting Repetition SchemeThe test sequence can be set to repeat itself for a cer

Pagina 104 - :INPut[1..n]:OFFSet

Setting Up Your Variable AttenuatorVariable Attenuator 41Creating a Test Sequence2. Under Sequence File, open a test sequence.ORCreate a sequence (ref

Pagina 105

Setting Up Your Variable Attenuator42 IQS-3150Managing Sequence FilesManaging Sequence FilesThe commands to open new or existing sequence files and to

Pagina 106 - :INPut[1..n]:OFFSet?

Setting Up Your Variable AttenuatorVariable Attenuator 43Managing Sequence Files³ Clicking the Open button opens the Open Sequence File dialog box fro

Pagina 107

Certification InformationVariable Attenuator vCertification InformationF.C.C. InformationElectronic test equipment is exempt from Part 15 compliance (

Pagina 108 - :INPut[1..n]:RATTenuation

Setting Up Your Variable Attenuator44 IQS-3150Saving and Recalling ConfigurationsSaving and Recalling ConfigurationsOnce you have set the IQS-3150 Var

Pagina 109

Setting Up Your Variable AttenuatorVariable Attenuator 45Saving and Recalling ConfigurationsTo recall a configuration:1. Select the Configuration func

Pagina 111 - :INPut[1..n]:RATTenuation?

Variable Attenuator 474 Operating the IQS-3150This section presents basic operations using the Variable Attenuator.Cleaning and Connecting Optical Fib

Pagina 112

Operating the IQS-315048 IQS-3150Installing the EXFO Universal Interface (EUI)Installing the EXFO Universal Interface (EUI)The EUI fixed baseplate is

Pagina 113 - :INPut[1..n]:REFerence

Operating the IQS-3150Variable Attenuator 49Nulling Electrical OffsetsThe nulling function is only available for the self-adjusting modules.To perform

Pagina 114

Operating the IQS-315050 IQS-3150Running an Attenuation SequenceRunning an Attenuation SequenceAttenuation sequences allow you to automate and ease yo

Pagina 115 - :INPut[1..n]:REFerence?

Operating the IQS-3150Variable Attenuator 51Running an Attenuation SequenceYou can interrupt a sequence at any time by clicking Pause; click Resume wh

Pagina 116

Operating the IQS-315052 IQS-3150Using the Power Tracking Function (Self-Adjusting Modules)Using the Power Tracking Function (Self-Adjusting Modules)T

Pagina 117 - :INPut[1..n]:WAVelength

Operating the IQS-3150Variable Attenuator 53Using the ShutterUsing the ShutterThe shutter is used to let light through the input or block it.To use th

Pagina 118

Certification Informationvi IQS-3150 Application of Council Directive(s): 73/23/EEC - The Low Voltage Directive 89/33

Pagina 119 - :INPut[1..n]:WAVelength?

Operating the IQS-315054 IQS-3150Using the ShutterShutter StatusThe status of the shutter is shown on the left of the data display.It is also shown on

Pagina 120

Variable Attenuator 555 Controlling Multiple Variable AttenuatorsWith your platform, you can set common parameters and simultaneously operate several

Pagina 121 - :OUTPut[1..n]:ALC[:STATe]

Controlling Multiple Variable Attenuators56 IQS-3150Selecting Modules to ControlSelecting Modules to ControlBefore you can modify the module parameter

Pagina 122 - :OUTPut[1..n]:ALC[:STATe]?

Controlling Multiple Variable AttenuatorsVariable Attenuator 57Controlling a Single IQS-3150 Variable AttenuatorOn the Instruments function tab, you c

Pagina 123 - :OUTPut[1..n]:APMode

Controlling Multiple Variable Attenuators58 IQS-3150Navigating and Closing Multiple Module WindowsNavigating and Closing Multiple Module WindowsWhen c

Pagina 124

Variable Attenuator 596 Monitoring Variable Attenuator ModulesWhen using your IQS-3150 Variable Attenuator module, either alone or with other modules

Pagina 125 - :OUTPut[1..n]:APMode?

Monitoring Variable Attenuator Modules60 IQS-3150Using Monitor WindowsTo select modules and display their monitor windows:1. On the Current Modules fu

Pagina 126 - :OUTPut[1..n]:DTOlerance

Monitoring Variable Attenuator ModulesVariable Attenuator 61Using Monitor Windows3. Click Start Monitor to apply your selection.IQS Manager will displ

Pagina 127

Monitoring Variable Attenuator Modules62 IQS-3150Using QuickToolsUsing QuickToolsWith QuickTools, you can fine-tune your module directly, while keepin

Pagina 128 - :OUTPut[1..n]:DTOlerance?

Monitoring Variable Attenuator ModulesVariable Attenuator 63Using QuickToolsTo c los e Q ui c kTo ol s:³ Click the Close button located at the top of

Pagina 129

Variable Attenuator 11 Introducing the IQS-3150 Variable AttenuatorThe IQS-3150 is a versatile variable attenuator. It can be used as a regular attenu

Pagina 131 - :OUTPut[1..n]:OFFSet

Variable Attenuator 657 Measuring Multimode Insertion LossOften, when using a multimode source (mostly LED sources), part of the optical energy is tra

Pagina 132

Measuring Multimode Insertion Loss66 IQS-3150There are several methods of eliminating these cladding modes:³ A cladding mode stripper (preferred metho

Pagina 133 - :OUTPut[1..n]:OFFSet?

Variable Attenuator 678 MaintenanceTo help ensure long, trouble-free operation:³ Always clean fiber-optic connectors before using them.³ Keep the unit

Pagina 134

Maintenance68 IQS-3150Cleaning Fixed ConnectorsCleaning Fixed ConnectorsRegular cleaning of connectors will help maintain optimum performance. Do not

Pagina 135 - :OUTPut[1..n]:POWer

MaintenanceVariable Attenuator 69Cleaning Fixed Connectors6. Moisten a cleaning tip (2.5 mm tip) with only one drop of isopropyl alcohol.7. Slowly ins

Pagina 136

Maintenance70 IQS-3150Cleaning EUI ConnectorsCleaning EUI ConnectorsRegular cleaning of EUI connectors will help maintain optimum performance. There i

Pagina 137 - :OUTPut[1..n]:POWer?

MaintenanceVariable Attenuator 71Cleaning EUI Connectors5. Repeat steps 3 to 4 with a dry cleaning tip.Note: Make sure you don’t touch the soft end of

Pagina 138

Maintenance72 IQS-3150Cleaning Detector PortsCleaning Detector PortsRegular cleaning of detectors will help maintain measurement accuracy.To clean det

Pagina 139 - :OUTPut[1..n]:REFerence

MaintenanceVariable Attenuator 73Homing the Variable Attenuator (User Calibration)Homing the Variable Attenuator (User Calibration)The variable attenu

Pagina 140

Introducing the IQS-3150 Variable Attenuator2 IQS-3150Models and OptionsThe IQS-3150 Variable Attenuator is available in four models:³ Standard single

Pagina 141 - :OUTPut[1..n]:REFerence?

Maintenance74 IQS-3150Recalibrating the UnitRecalibrating the UnitManufacturing and service center calibrations are based on the ISO/IEC 17025 Standar

Pagina 142

MaintenanceVariable Attenuator 75Recycling and Disposal (Applies to European Union Only)Recycling and Disposal (Applies to European Union Only)³ Unles

Pagina 144

Variable Attenuator 779 TroubleshootingViewing Online DocumentationAn online version of the IQS-3150 Variable Attenuator user guide is conveniently av

Pagina 145 - :OUTPut[1..n]:RPOWer?

Troubleshooting78 IQS-3150Contacting the Technical Support GroupContacting the Technical Support GroupTo obtain after-sales service or technical suppo

Pagina 146

TroubleshootingVariable Attenuator 79Contacting the Technical Support GroupYou may also be requested to provide software and module version numbers. T

Pagina 147 - :OUTPut[1..n][:STATe]

Troubleshooting80 IQS-3150TransportationTransportationMaintain a temperature range within specifications when transporting the unit. Transportation da

Pagina 148 - :OUTPut[1..n][:STATe]?

Variable Attenuator 8110 WarrantyGeneral InformationEXFO Electro-Optical Engineering Inc. (EXFO) warrants this equipment against defects in material a

Pagina 149 - Variable Attenuator 143

Warranty82 IQS-3150ExclusionsExclusionsEXFO reserves the right to make changes in the design or construction of any of its products at any time withou

Pagina 150 - 144 IQS-3150

WarrantyVariable Attenuator 83Service and RepairsService and RepairsEXFO commits to providing product service and repair for five years following the

Pagina 151 - :SNUMber?

Introducing the IQS-3150 Variable AttenuatorVariable Attenuator 3Typical ApplicationsTypical ApplicationsThe IQS-3150 Variable Attenuator can be used

Pagina 152 - :STATus?

Warranty84 IQS-3150EXFO Service Centers WorldwideEXFO Service Centers WorldwideIf your product requires servicing, contact your nearest authorized ser

Pagina 153 - Variable Attenuator 147

Variable Attenuator 85A Technical SpecificationsIMPORTANTThe following technical specifications can change without notice. The information presented i

Pagina 154 - on the internal power meter

Technical Specifications86 IQS-3150GENERAL SPECIFICATIONSSize (H X W X D) 125 mm X 36 mm X 282 mm (4 15/16in X 1 7/16in X 11 1/8in)Weight 0.7 kg (1.6

Pagina 155 - CONDition?

Variable Attenuator 87B SCPI Command ReferenceThis appendix presents detailed information on the commands and queries supplied with your IQS-3150 Vari

Pagina 156

SCPI Command Reference88 IQS-3150Quick Reference Command TreeQuick Reference Command TreeCommand Parameter(s) P.CALibration[1..n]ZERO 90CONTrol[1..n]

Pagina 157 - Variable Attenuator 151

SCPI Command ReferenceVariable Attenuator 89Quick Reference Command TreeREFerence <Reference[<wsp>DBM]>|MAXimum|MINimum|DEFault133REFerenc

Pagina 158 - ... 48

SCPI Command Reference90 IQS-3150Product-Specific Commands—DescriptionProduct-Specific Commands—Description:CALibration[1..n]:ZERODescription Returns

Pagina 159 - Variable Attenuator 153

SCPI Command ReferenceVariable Attenuator 91Product-Specific Commands—Description:CONTrol[1..n]:MODEDescription This command selects the attenuator&ap

Pagina 160

SCPI Command Reference92 IQS-3150Product-Specific Commands—Description:CONTrol[1..n]:MODE?Description This query returns the attenuator's control

Pagina 161

SCPI Command ReferenceVariable Attenuator 93Product-Specific Commands—Description:INPut[1..n]:ARESolution?Description This query returns the smallest