Technical Specifications 33
Model FOT-302 FOT-302X
Power meter port
b
Ge GeX
Power range (dBm)
c
10 to —60 26 to —50
Range displayed (dBm) Down to —65 Down to —50
Number of calibrated wavelengths
d
10 10
Power uncertainty
e
± 5 % ± 1 nW ± 5 % ± 10 nW
Resolution (dB) 0.01
f
0.01
g
Automatic offset nulling
h
Yes Yes
Warmup time (s)
h, i
0 0
Display units dB/dBm/W dB/dBm/W
Automatic wavelength recognition
j
Yes Yes
Screen refresh rate (Hz) 3 3
Tone detection (Hz) 270, 1 k, 2 k 270, 1 k, 2 k
Battery life (hours) (typical) 260 260
Warranty and recommended calibration interval (years) 3 3
Model
k
23BL 12D 235BL
Central wavelength (nm) 1310 ± 20 850 ± 25/ 1310 ± 20
1550 ± 20 1300 + 50/—10 1490 ± 10
1550 ± 20
Spectral width (nm)
l
) 5 50/135 ) 5
Output power (dBm) * 1/* 1
* —20/* —20 (62.5/125 +m) * 1/* —4.5/*—3
Power stability (dB)
m
8 hours ± 0.1 ± 0.1 ± 0.1
Battery life (hours)
n
120 120 120
Automatic wavelength recognition Yes Yes Yes
Tone generation (Hz) 270, 1 k, 2 k 270, 1 k, 2 k 270, 1 k, 2 k
Warranty and recommended calibration interval (years) 3 3 3
SPECIFICATIONS
a
GENERAL SPECIFICATIONS
Size (H x W x D) 185 mm x 100 mm x 55 mm (7
1
/
4
in x 4 in x 2
1
/
8
in)
Weight 0.4 kg (0.9 lb)
Temperature operating —10 °C to 50 °C (14 °F to 122 °F)
storage —40 °C to 70 °C (—40 °F to 158 °F)
Relative humidity 0 % to 95 % non-condensing
STANDARD ACCESSORIES
User guide, Certificate of Calibration, instrument stickers in six languages,
AC adapter, EUI-XX, connector adapter (FOA-XX), three AA batteries, wrist
strap, alcohol cleaning pads.
SAFETY
21 CFR 1040.10 and IEC 60825-1:1993+A1:1997+A2:2001:
CLASS 1 LASER PRODUCT CLASS 1 LED PRODUCT
NOTES
a. Guaranteed unless otherwise specified.
b. All specifications valid at 1550 nm and 23 °C ± 1 °C, with
an FC connector.
c. In CW mode; sensitivity defined as 6 x rms noise level.
d. Wavelengths: 830 nm, 850 nm, 980 nm, 1300 nm, 1310
nm, 1450 nm, 1490 nm,
1550 nm, 1590 nm and 1625 nm.
e. Traceable to national standards; FOT-302X: up to 20 dBm.
f. From 10 dBm to
50 dBm.
g. From 26 dBm to
35 dBm.
h. Power of > —40 dBm for FOT-302, and of >
25 dBm for
FOT-302X.
i. For ± 0.05 dB, from 18 °C to 28 °C.
j. At 850 nm, 1300 nm, 1310 nm, 1490 nm, 1550 nm and
1625 nm; power > —50 dBm for FOT-302, and
> —40 dBm (typical) for FOT-302X.
k. All specifications valid at 23 °C ± 1 °C, with an FC
connector.
l. rms for lasers and —3 dB width for LEDs; typical values for
LEDs.
m. After 15 minutes warmup; expressed as ± half the difference
between the maximum and minimum values measured during
the period, with an APC connector on the power meter.
n. Typical autonomy in Auto mode.
Vista de pagina 37
1 2 ... 33 34 35 36 37 38 39 40

Comentarios a estos manuales