Exfo FLS-5800 CD/PMD Analyzer Source Manual de usuario Pagina 97

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 106
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 96
Instrucciones de control remoto
Fuente de analizador de CD/PMD 91
Instrucciones específicas del producto: Descripción
SOURce:POWer[:STATe]
Description This command turns on or off the source. When the source is
on, the red LED on the front of the module illuminates.
Syntax SOUR:POW[:STAT]<wsp><boolean>|ON|OFF
Parameters A boolean parameter:
“1”–turns the source on
“0”–turns the source off
Example SOUR:POW:STAT 1 (turns the source on.)
SOURCE:POWER 0 (turns the source off.)
SOURce:POWer:STATe?
Description This query returns a value indicating the state of the optical
source (on or off).
Syntax SOUR:POW:STAT?
Response A boolean value:
“1”–the source is on
“0”–the source is off
Example SOUR:POW:STAT? 1
Vista de pagina 96
1 2 ... 92 93 94 95 96 97 98 99 100 101 102 ... 105 106

Comentarios a estos manuales

Sin comentarios