Remote Control Commands
CD/PMD Analyzer Source 83
Product-Specific Commands-Description
SOURce:POWer[:STATe]
Description This command turns on or off the source. When the source is
on, the red LED on the front of the module illuminates.
Syntax SOUR:POW[:STAT]<wsp><boolean>|ON|OFF
Parameters A boolean parameter:
“1”–turns the source on
“0”–turns the source off
Example SOUR:POW:STAT 1 (turns the source on.)
SOURCE:POWER 0 (turns the source off.)
SOURce:POWer:STATe?
Description This query returns a value indicating the state of the optical
source (on or off).
Syntax SOUR:POW:STAT?
Response A boolean value:
“1”–the source is on
“0”–the source is off
Example SOUR:POW:STAT? 1