SCPI-Befehlsreferenz
OTDR 453
Product-Specific Commands—Description
:SOURce[1..n]:VFLocator:POWer:STATe
Description This command turns the visual fault locator
(VFL) on or off.
*RST sets the visual fault locator to OFF.
Syntax :SOURce[1..n]:VFLocator:POWer:STATe<wsp>
<State>
Parameter(s) State:
The program data syntax for <State> is defined
as a <Boolean Program Data> element. The
<State> special forms ON and OFF are accepted
on input for increased readability. ON
corresponds to 1 and OFF corresponds to 0.
New power state of the VFL.
Example(s) SOUR:VFL:AM:INT:FREQ 1
SOUR:VFL:POW:STAT:TIME 60
SOUR:VFL:AM:STAT ON
SOUR:VFL:POW:STAT ON
See Also SOURce[1..n]:VFLocator:POWer:STATe?
SOURce[1..n]:VFLocator:POWer:STATe:TIME
SOURce[1..n]:VFLocator:AM:STATe?
SOURce[1..n]:VFLocator:AM:INTernal:FREQuenc
y