
SCPI-Befehlsreferenz
OTDR 409
Product-Specific Commands—Description
:FETCh[1..n]:ASETting:PULSe?
Description This query returns the pulse found after an
initiate (INIT) command. Note that acquisition
mode (CONF:ACQ:MODE) must be set to
ASETting.
Since *RST clears the pulse value, the returned
value will be 0.
Syntax :FETCh[1..n]:ASETting:PULSe?
Parameter(s) None
Response Syntax <Pulse>
Response(s) Pulse:
The response data syntax for <Pulse> is defined
as a <NR3 NUMERIC RESPONSE DATA>
element.
Returns the pulse, in meters.
Example(s) CONF:ACQ:MODE ASET
INIT
INIT:STAT? Returns 0 when acquisition is
complete.
FETC:ASET:PULS? Ex.: Returns 1E–8
See Also CONFigure[1..n]:ACQuisition:RANGe?
CONFigure[1..n]:ACQuisition:PULSe?
CONFigure[1..n]:ACQuisition:DURation?
CONFigure[1..n]:ACQuisition:WAVelength:LIST?
CONFigure[1..n]:ACQuisition:RANGe:LIST?
CONFigure[1..n]:ACQuisition:PULSe:LIST?
CONFigure[1..n]:ACQuisition
Comentarios a estos manuales