SCPI Command Reference
Optical Spectrum Analyzer 507
Product-Specific Commands—Description
:SENSe[1..n]:WAVelength[:RANGe]:
STOP?
Description This query returns the start wavelength for
acquisition range.
At *RST, this value is set to 1565.0 nm (C-band
upper limit).
Syntax :SENSe[1..n]:WAVelength[:RANGe]:STOP?
Parameter(s) None
Response Syntax <Stop>
Response(s) Stop:
The response data syntax for <Stop> is defined
as a <NR3 NUMERIC RESPONSE DATA>
element.
Maximum wavelength for acquisition range (in m
or Hz).
Example(s) SENS:WAV:RANG:STOP?
See Also SENSe[1..n]:WAVelength:RANGe:STOP
SENSe[1..n]:WAVelength:RANGe:STARt?