SCPI Command Reference
514 FTB-5240S/5240BP
Product-Specific Commands—Description
:TRACe[1..n]:INFormation?
Description This query returns all available information on a
trace.
*RST has no effect on traces in memory.
Syntax :TRACe[1..n]:INFormation?<wsp>TRCMAIN
Parameter(s) TracePosition:
The program data syntax for the first parameter is
defined as a <CHARACTER PROGRAM DATA>
element. The allowed <CHARACTER PROGRAM
DATA> element for this parameter is TRCMAIN.
Selects the Trace Index.
Response Syntax <TraceInfo>
Response(s) TraceInfo:
The response data syntax for <TraceInfo> is
defined as a <DEFINITE LENGTH ARBITRARY
BLOCK RESPONSE DATA> element.
Selected trace information A,B,C,D,E,F,G format,
where:
Vista de pagina 521