Managing Results
Optical Spectrum Analyzer 381
Managing EDFA Test Results
³ Channel index: indicates the channel number. Channel 1
corresponds to the first channel found.
³ Noise figure: indicates the EDFA’s noise figure measured for the
selected channel.
³ S%: indicates the current output power according to the
measured output power (Output Signal Power / [Output Signal
Power + PA SE]).
³ PASE: indicates the power of the spontaneous emission amplified
by EDFA.
³ PSSE: indicates the power of the spontaneous emission of the
source.
4. Under Analysis parameters pane, you can view the following results:
³ Channel center: indicates the channel center for the selected
channel.
³ Channel width: indicates the limit within which the power values
will be considered in the channel.