Exfo FTB -5240S/5240BP Optical Spectrum Analyzer for FT Manual de usuario Pagina 313

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 537
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 312
Managing Results
Optical Spectrum Analyzer 305
Managing WDM Test Results
³ Total power in scan range: indicates the total power for the
selected scan range.
³ Channel count: indicates the number of channels detected after
an acquisition for the selected scan range.
³ Empty channel count: indicates the number of empty channels
from the channel list.
³ Peak detection level: indicates the minimum power level from
where the peak can be considered as a signal.
³ RBW for OSNR: indicates the resolution bandwidth for the
selected OSNR value.
³ Wavelength offset: indicates the offset value applied on the
wavelength.
³ Power offset: indicates the offset value applied on the power.
In addition, you can also view the global pass/fail status, provided the
thresholds are activated in the Global Result Thresholds tab in the
Analysis Setup window. If the thresholds are disabled, the Global
pass/fail status pane will display Not Active. If the thresholds are
activated, the Global pass/fail status pane will display either a Pass or Fail
based on the global results.
Note: To edit global parameters, see Changing Active Trace Analysis Parameters
and Re-analyzing on page 306.
Vista de pagina 312
1 2 ... 308 309 310 311 312 313 314 315 316 317 318 ... 536 537

Comentarios a estos manuales

Sin comentarios