OSA/OCA 241
A Especificaciones técnicas
IMPORTANTE
Las siguientes especificaciones técnicas pueden cambiar sin previo
aviso. La información contenida en esta sección se proporciona
únicamente como referencia. Si desea obtener las especificaciones
técnicas más recientes del producto, visite la página web de EXFO
en www.exfo.com.
POWER MEASUREMENT
Dynamic range (dBm) (per channel)
b
-65
d
to 23 dB
Maximum total safe power (dBm) 29
Absolute power uncertainty (dB)
h
±0.6 (0.4 typical)
Power repeatability 2m (dB)
d, g
±0.1
SPECTRAL MEASUREMENT
Wavelength range (nm) 1250 to 1650
Wavelength uncertainty (nm)
b
±0.05
±0.02
c, d
Reference Internal
e
Resolution bandwidth (FWHM)
f
(nm) 0.10
b, d
Wavelength repeatability 2m (nm) ±0.005
g
Analysis modes WDM and drift (FTB-5230S) / Optical Channel Analyzer mode (FTB-5230S-OCA)
GENERAL SPECIFICATIONS
Temperature
operating
storage
0 °C to 40 °C (32 °F to 104 °F)
—20 °C to 50 °C (—4 °F to 120 °F)
Connectors
EI (EXFO UPC Universal Interface)
EA (EXFO APC Universal Interface)
Size (H x W x D) FTB-5230S module 96 mm x 51 mm x 260 mm (3 ¾ in x 2 in x 10 ¼ in)
Weight FTB-5230S module 1. 5 kg (3.3 lb)
UPGRADES
WDM: enables upgrade from FTB-5230S-OCA
to FTB-5230S (software option)
LASER SAFETY
21 CFR 1040.10 AND IEC 60825-1
CLASS 1 LASER PRODUCT
OPTICAL MEASUREMENT
Optical rejection ratio at 1550 nm (dB)
at 0.2 nm (25 GHz)
at 0.4 nm (50 GHz)
31 (35 typical)
40 (45 typical)
Channel spacing 33 to 200 GHz CWDM
PDL at 1550 nm (dB) ±0.1
d
ORL (dB) >37
Measurement time (s)
d, i
(includes scanning, analysis and display)
1 (with the FTB-500 Platform)
Notes
a. All specifications are for a temperature of 23 °C ± 2 °C with an FC/UPC connector unless
otherwise specified, after warm-up.
b. From 1520 to 1600 nm.
c. For FTB-5230S only. After user calibration in the same test session within 10 nm from each
calibration point.
d. Typical.
e. Integrated and wavelength-independent self-adjustment.
f. Full width at half maximum.
g. Over one minute in continuous acquisition mode.
h. At 1550 nm, -10 dBm input.
i. 40 nm span.