Exfo FLS-2200 Broadband Source Manual de usuario Pagina 148

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 183
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 147
IEEE 488.2 and Specific Command Reference
140 FLS-2200
Product-Specific Commands—Description
:SOURce[1..n]:CURRent[:LEVel]
[:IMMediate][:AMPLitude]
Description This command sets the source's current level.
This value can be set even if the source is off.
At *RST, the current level is device-dependent.
Syntax :SOURce[1..n]:CURRent[:LEVel][:IMMediate][:A
MPLitude]<wsp><Current[<wsp>A]>|MAXim
um|MINimum
Parameter(s) Current:
The program data syntax for <Current> is
defined as a <numeric_value> element
followed by an optional <SUFFIX PROGRAM
DATA> element. The allowed <SUFFIX
PROGRAM DATA> element is A. The <Current>
special forms MINimum and MAXimum are
accepted on input.
MINimum allows to set the instrument to the
smallest supported value.
MAXimum allows to set the instrument to the
greatest supported value.
The <Current> parameter is the new source
current drive, in amperes (A).
Vista de pagina 147
1 2 ... 143 144 145 146 147 148 149 150 151 152 153 ... 182 183

Comentarios a estos manuales