148 AXS-200/630
DATA ANALYSIS MODE
Physical-layer support VDSL2
ADSL1/2/2+
Ethernet 10/100
Encapsulation methods PPPoE (RFC 2516), RFC 2684 supporting bridged Ethernet (IPoE), IPoA (RFC 1577), PPPoA/LLC
and PPPoA/VC-MUX (RFC 2364)
Operation Terminate and Pass Through
Login format Username and password using PAP or CHAP
IP connectivity support DNS, DHCP client/server, NAT, VLAN
Ping Pings another device on the network
Ping device: Gateway, destination IP address or URL
Number of pings: 1 to 99
Packet size: 32 to 1500 bytes (32 is default)
Results: Indicate packet size, packets sent/received, minimum/average/maximum round-trip times
in milliseconds (ms)
Traceroute Determines the path used to reach device on the network
Timeout: In seconds
Time to live (TTL): Default is 100 ms, maximum is 5 s
Packet size: 32 bytes
Number of hops: 1 to 30 (default is 30)
Results: Indicate IP address of hop and round-trip time in milliseconds (ms)
HTTP speed test Downloads a Web page and indicates speed of download
Address: IP or URL
Protocol: HTTP
Results: Time, speed in kbit/s
FTP speed test Displays speed to upload and/or download a file
Address: IP or URL
Protocol: FTP
Results: Time, speed in kbit/s