Automating or Remotely Controlling the System
190 IQS-12004B
GPIB Specific Commands—Quick Reference
IL -- This command performs an IL
reference acquisition.
REF IL? (0 | 1) The response to this query indicates
if an IL reference acquisition has
been performed.
ILPD -- This command performs a PDL
reference acquisition.
ILPD? (0 | 1) The response to this query indicates
if a PDL reference acquisition has
been performed.
TEST AVAT? (1 | 2 | 3 | 4 | 5 | 6) The response to this query indicates
which test can be performed
according to the modules detected,
the calibration, and the reference.
CLR -- This command clears the last IL,
PDL, and ORL results.
ACQ IL -- This command performs an IL
acquisition on the DUT.
ILPD -- This command performs a PDL
acquisition on the DUT.
ORL -- This command performs an ORL
acquisition on the DUT.
SYNC <99999> This command performs a
synchronized power acquisition on
all IQS-1600 power meters.
ACQ STOP -- This command stops the current
acquisition on the DUT.
STAT ERR? (Error Number,
“Error description”)
The response to this query indicates
the last error and provides a
description.
Command Parameter/Response Description