Exfo IQS-12004B DWDM Passive Component Test System for Manual de usuario Pagina 128

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 437
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 127
Referencing the System with MultiPath Testing Option
116 IQS-12004B
Performing Reference Measurements
Performing Reference Measurements
An IL reference is performed to compensate for any loss or spectral
non-uniformity of the launch-fiber connection to the WRM output port.
To perform an IL reference:
1. Connect output port 1 of the switch to channel 1 of the power meter.
2. Perform an IL reference measurement.
3. Repeat the above steps for each of the remaining switch output ports
by connecting the switch port to channel 1 of the power meter.
An IL/PDL reference is performed to compensate for any loss or spectral
non-uniformity of the launch fiber connection to the WRM output port.
To perform an IL/PDL reference:
Note: References are valid as long as rigid and semi-rigid patchcords do not
move.
1. Connect output port 1 of the switch to channel 1 of the power meter.
2. Perform an IL/PDL reference measurement.
3. Repeat the above steps for each of the remaining switch output ports
by connecting the switch port to channel 1 of the power meter.
The return-loss zero measurement provides a zero-reflection reference to
the WRM return loss detector. This reference will be used to compensate
for internal WRM parasitic reflections as well as parasitic reflections
originating at the WRM output/launch-fiber connector.
To perform an ORL zero-reference measurement:
1. Connect the launch fiber to output port 1 of the switch.
2. Mandrel the launch fiber.
3. Perform the measurement.
4. Repeat the preceding steps for each switch port.
Vista de pagina 127
1 2 ... 123 124 125 126 127 128 129 130 131 132 133 ... 436 437

Comentarios a estos manuales

Sin comentarios