Exfo FTB-9100 for FTB-500 Optical Switch Manual de usuario Pagina 51

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 72
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 50
SCPI Command Reference
Optical Switch 45
Product-Specific Commands—Description
:ROUTe[1..n]:SCAN
Description This command sets the switch to a specific
channel.
On *RST, the selected output channel depends
on the switch configuration:
a) For 1xN switch configuration: Output 1.
b) For 2x4 switch configuration: no continuity for
both Input 1 and Input 2.
c) For 2x2 switch configuration: BYPASS state.
Syntax :ROUTe[1..n]:SCAN<wsp><Position>
Parameter(s) Position:
The program data syntax for <Position> is
defined as a <DECIMAL NUMERIC PROGRAM
DATA> element.
The <Position> is a number from 0 to 32,
depending on switch configuration as follows:
a) For a 1xN switch configuration, a number from
1 to 32, depending on the number of output
channels.
b) For a 2x4 switch, a number from 0 to 5, where
0, means no continuity for both Input 1 and
Input 2.
1, means no continuity for Input 1, and Input 2
is connected to Output 1.
Vista de pagina 50