Exfo FTB-700 for FTB-1 Manual de usuario Pagina 304

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 329
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 303
Description of Event Types
296 FTB-700 Series
End of Analysis
End of Analysis
This event indicates that the pulse width used did not provide enough
dynamic range to get to the end of the fiber.
The analysis ended before reaching the end of the fiber because the
signal-to-noise ratio was too low.
The pulse width should therefore be increased so the signal reaches
the end of the fiber with a sufficient signal-to-noise ratio.
There is no loss or reflectance specified for end-of-analysis events.
Reflected power
(dB)
Distance
(km)
Vista de pagina 303
1 2 ... 299 300 301 302 303 304 305 306 307 308 309 ... 328 329

Comentarios a estos manuales

Sin comentarios