Exfo FTB-5800 Chromatic Dispersion Analyzer for FTB-400 Manual de usuario Pagina 63

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 103
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 62
Managing Results
Chromatic Dispersion Analyzer 57
Establishing Thresholds
The Slope 0 Threshold is the slope of the dispersion curve evaluated at
the points where the dispersion is null (lambda 0). In the Results Graph
tab, you will have four possible results:
³ Pass if the slope at lambda 0 is less than or equal to the threshold.
³ Fail if it is greater than the threshold.
³ Fit if your slope at lambda 0 did not respect FOTP standard 169. This
standard requires other points within 100 nm; otherwise, the slope at
lambda 0 is not reliable and can only be used as a reference.
³ --- if you did not select any thresholds.
Vista de pagina 62
1 2 ... 58 59 60 61 62 63 64 65 66 67 68 ... 102 103

Comentarios a estos manuales

Sin comentarios