Exfo FTB-5700 Single Ended Dispersion Analyzer for FTB- Manual de usuario Pagina 77

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 94
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 76
Troubleshooting
Single-Ended Dispersion Analyzer 71
Solving Common Problems
There are not enough
valid points to compute
the model fit due to
detector saturation
conditions in the
selected range.
³ Fiber end is too
reflective.
³ Unsuitable FUT.
The FUT must not
cut the 1550 nm
wavelength.
³ If you have put a reflective
termination at the end of the
fiber, remove it.
There are not enough
valid points to compute
the model fit due to a
combination of detector
saturation and low
power conditions in the
selected range.
³ Fiber end is too
reflective.
³ Unsuitable selected
wavelength range
for the measured
FUT (out of the FUT
bandwidth).
³ Unsuitable FUT.
The FUT must not
cut the 1550 nm
wavelength.
³ If you have put a reflective
termination at the end of the
fiber, remove it.
³ Select a wavelength range that
suits the FUT and the instrument
range. You can set the
wavelength range in the
Acquisition tab of the Setup
window.
Message Possible cause Solution
Vista de pagina 76
1 2 ... 72 73 74 75 76 77 78 79 80 81 82 ... 93 94

Comentarios a estos manuales

Sin comentarios