Exfo FTB-5500B PMD Analyzer Manual de usuario Pagina 49

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 121
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 48
Operating the Polarization Mode Dispersion Analyzer
Polarization Mode Dispersion Analyzer 41
Performing a Null Measurement
Performing a Null Measurement
This feature allows you to perform an offset nulling of your FTB-5500B
Polarization Mode Dispersion Analyzer. During this operation, no light
should reach the detector port of the Polarization Mode Dispersion
Analyzer.
The very first time you insert an FTB-5500B Polarization Mode Dispersion
Analyzer into the FTB-500, the application will prompt you for a nulling.
EXFO recommends that you perform a null measurement just before
verifying the input power level to ensure the accuracy of the displayed
level.
To perform a null measurement:
1. From the main window, click the Parameters tab. Click Nulling and
follow the on-screen instructions.
2. When nulling is complete, click Close.
Vista de pagina 48
1 2 ... 44 45 46 47 48 49 50 51 52 53 54 ... 120 121

Comentarios a estos manuales

Sin comentarios