Exfo FTB -5240S/5240BP Optical Spectrum Analyzer for FT Manual de usuario Pagina 537

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 653
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 536
SCPI-Befehlsreferenz
OSA 529
Product-Specific Commands—Description
:CALCulate[1..n][:WDM]:THReshold?
Description This query returns a value indicating either the
current or the minimum/maximum WDM
analysis absolute power threshold for peak
detection.
At *RST, this value is set to -45.0 dBm.
Syntax :CALCulate[1..n][:WDM]:THReshold?[<wsp>M
AXimum|MINimum|DEFault]
Parameter(s) Parameter 1:
The program data syntax for the first parameter is
defined as a <CHARACTER PROGRAM DATA>
element. The allowed <CHARACTER PROGRAM
DATA> elements for this parameter are:
MAXimum|MINimum|DEFault.
MINimum is used to retrieve the instrument's
smallest supported value.
MAXimum is used to retrieve the instrument's
greatest supported value.
DEFault is used to retrieve the instrument's
default value.
Response Syntax <Threshold>
Vista de pagina 536
1 2 ... 532 533 534 535 536 537 538 539 540 541 542 ... 652 653

Comentarios a estos manuales