SCPI-Befehlsreferenz
OSA 525
Product-Specific Commands—Description
:CALCulate[1..n][:WDM]:STATe
Description This command controls activation of WDM
analysis.
At *RST, WDM analysis is set to on (enabled).
Syntax :CALCulate[1..n][:WDM]:STATe<wsp><Auto>
Parameter(s) Auto:
The program data syntax for <Auto> is defined
as a <Boolean Program Data> element. The
<Auto> special forms ON and OFF are accepted
on input for increased readability. ON
corresponds to 1 and OFF corresponds to 0.
The <Auto> parameter corresponds to the new
state of WDM analysis.
0 or OFF: WDM analysis is disabled.
1 or ON: WDM analysis is enabled.
Example(s) CALC:WDM:STAT ON
CALC:WDM:STAT? Returns 1 (WDM analysis
enabled)
Notes WDM analysis cannot be disabled: OFF (0) value
is valid for queries only.
It is possible to query acquired trace data only if
active measurement analysis mode is WDM.
See Also :CALCulate[1..n][:WDM]:STATe?