SCPI-Befehlsreferenz
562 FTB-5240S/S-P/BP
Product-Specific Commands—Description
:SENSe[1..n]:CORRection:OFFSet
[:MAGNitude]
Description This command sets the power offset that is
added to every point measured by the
instrument.
At *RST, this value is set to 0.0 dB.
Syntax :SENSe[1..n]:CORRection:OFFSet[:MAGNitude]
<wsp><Offset[<wsp>DB|W/W|PCT]>|MAXi
mum|MINimum|DEFault
Parameter(s) Offset:
The program data syntax for <Offset> is defined
as a <numeric_value> element followed by an
optional <SUFFIX PROGRAM DATA> element.
The allowed <SUFFIX PROGRAM DATA>
elements are: DB|W/W|PCT. The <Offset>
special forms MINimum, MAXimum and DEFault
are accepted on input.
MINimum allows to set the instrument to the
smallest supported value.
MAXimum allows to set the instrument to the
greatest supported value.