Performing Automated IL/ORL/Length Measurements (FASTEST)
48 FOT-930
Setting Up the FASTEST
Setting Up the FASTEST
You can configure the FASTEST on the master unit only. The remote unit will
automatically adapt to these parameters (F
ASTEST parameters on remote
unit are ignored).
Setup includes the following elements:
Port and DUT (fiber) type: multimode
F
ASTEST will not include ORL.
Compatibility: select FOT-930 for fast, two-
or three-wavelength testing including ORL (it
requires two FOT-930/FTB-3930). Use
FOT-920 when other unit is an FOT-920 or
FTB-3920.
Length units: affects thresholds and fiber length only.
Mode/wavelengths: select one or more wavelengths for the FASTEST.
Depending on your choice, the F
ASTEST will include loss and/or ORL
measurements. Selecting FTTx Custom or FTTx All allows you to
define upstream and downstream wavelengths.
Automatic saving of results: select where results are automatically
stored (remote saving not available with FOT-920). Fiber name
automatically increments after each F
ASTEST.
Unit location: in FTTx mode, you specify if the master unit is closer to
the CO (or to the premises) than the remote unit.
To set up the F
ASTEST:
1. Press Menu, then select Setup > FasTesT.
2. Select the F
ASTEST parameters.
(standard wavelengths)
(FTTx wavelengths)