Common Tab
Ethernet and Fibre Channel Application 361
Performance Monitoring (PM)
SES (Severely Errored Second)
For G.821: Gives the number of seconds within which a bit error ratio
is 10
-3
, or during which Loss Of Signal (LOS), Link down, or Pattern
Loss is detected.
For G.826: Gives the number of seconds within which EB count is
SES threshold (30%), or at least one NE defect occurred.
BBE (Background Block Error) (G.826 only): Gives the count of Errored
Block not occurring as part of a SES. A Block corresponds to a
complete Ethernet/Fibre Channel frame.
AS (Available Second): Gives the count of the seconds corresponding
to the periods of time within which no SES errors occurred for at least
10 consecutive seconds.