Exfo PM-1600 High-Speed Power Meter Manual de usuario Pagina 126

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 162
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 125
Remote Control (SCPI) Commands
116 PM-1600
General Commands
SYSTem:ERRor?
Description This command returns the next error on the list. When an error
is generated, an error number is sent to the error list. The error
list is accessed with the SYST:ERR? query. If the list contains 20
errors and a new error occurs, the first error will be erased.
Syntax SYST:ERR?
Response See error list and descriptions in Error Messages on page 80.
SYSTem:VERSion?
Description This query returns the current system software version.
Syntax SYST:VERS?
Response “EXFO E.O. Engineering PM-1600 Vxx.xx”, where xx.xx is the
current product version.
Note The *IDN? and SYST:VERS? queries are equivalent. They give the
same result.
*TRG
Description This command is not supported
Syntax *TRG
Remote Control (SCPI) Commands
116 PM-1600
General Commands
SYSTem:ERRor?
Description This command returns the next error on the list. When an error
is generated, an error number is sent to the error list. The error
list is accessed with the SYST:ERR? query. If the list contains 20
errors and a new error occurs, the first error will be erased.
Syntax SYST:ERR?
Response See error list and descriptions in Error Messages on page 80.
SYSTem:VERSion?
Description This query returns the current system software version.
Syntax SYST:VERS?
Response “EXFO E.O. Engineering PM-1600 Vxx.xx”, where xx.xx is the
current product version.
Note The *IDN? and SYST:VERS? queries are equivalent. They give the
same result.
*TRG
Description This command is not supported
Syntax *TRG