SmartR™ Features
128 MAX-635
Pair Detective
Pair Detective Threshold Settings
The Threshold Settings page allows you to set the
Pair Detective Threshold Settings as follow:
Foreign DC Voltage sets the maximum foreign DC
voltage value.
POTS DC Voltage sets the minimum working
POTS DC voltage value.
POTS DC Current sets the minimum working
POTS DC current value.
Capacitive Balance sets the minimum capacitive
balance threshold as a percentage.
Good VF/AC Balance sets the good VF/AC balance threshold.
Marginal VF/AC Balance sets the marginal VF/AC balance threshold.
Isolation sets the isolation resistance threshold.
Type is a read-only field displaying the threshold limit: either Minimum
or Maximum.
Enabled is a read-only field and always Yes for all thresholds.
To set threshold values:
1. Press the up/down arrow keys to highlight the desired parameter.
2. Press on a value to open the alphanumeric editor screen and use
the navigation keys to scroll through.
3. Press to select each value and use the function keys to complete
your selection.
4. Press to confirm and close your selections.