Analyzing Traces and Events
154 MAX-700
Setting the Analysis Detection Thresholds
To set the analysis detection thresholds:
1. From the button bar, press OTDR Setup.
2. From the OTDR Setup dialog box, select the Analysis tab.
3. Under Analysis parameters, set the parameters.
Enter the desired values in the appropriate boxes.
OR
Select the default settings by pressing Default.
4. Press Exit OTDR Setup.
The analysis detection thresholds you have just set are applied to all newly
acquired traces.
Note: Analysis thresholds are only saved in the trace during analysis. For traces
acquired, but not yet analyzed, you can change the analysis detection
thresholds in the OTDR test application before performing the analysis.