SCPI Command Reference
170 IQS-1700
Product-Specific Commands—Description
:SENSe[1..n]:POWer:WAVelength?
Description This query returns the currently selected
wavelength on the specified channel. If no
channel was specified, the default value used is
1.
Syntax :SENSe[1..n]:POWer:WAVelength?[<wsp>MINi
mum|MAXimum|DEFault]
Parameter(s) Parameter 1:
The program data syntax for the first parameter is
defined as a <CHARACTER PROGRAM DATA>
element. The allowed <CHARACTER PROGRAM
DATA> elements for this parameter are:
MINimum|MAXimum|DEFault.
MINimum is used to retrieve the instrument's
smallest supported value.
MAXimum is used to retrieve the instrument's
greatest supported value.
DEFault is used to retrieve the instrument's
default value.
Response Syntax <Wavelength>
Response(s) Wavelength:
The response data syntax for <Wavelength> is
defined as a <NR3 NUMERIC RESPONSE DATA>
element.
This query returns the current wavelength in
meters for the specified channel.
Example(s) SENS:POW:WAV?
See Also SENSe:POWer:WAVelength