COPPER SPECIFICATIONS
a, b, c
Transmitter characteristics
Frequency range (200 Hz to 20 kHz)
Frequency resolution 1 Hz steps
Frequency uncertainty (accuracy) ±(50 ppm + 1 Hz)
Level range –20 dBm to 0 dBm at 600 1
Level resolution 0.1 dB
Level uncertainty (accuracy) ±1 dB
Frequency range (20 kHz to 2.2 MHz)
Frequency resolution 1 kHz steps
Frequency uncertainty (accuracy) ±(50 ppm + 100 Hz)
Level range –20 dBm to 0 dBm at 100 1
Level resolution 0.1 dB
Level uncertainty (accuracy) ±1 dB
Frequency range (2.2 MHz to 17 MHz)
Frequency resolution 1 kHz steps
Frequency uncertainty (accuracy) ±(50 ppm + 100 Hz)
Level range –20 dBm to 0 dBm at 100 1
Level resolution 0.1 dB
Level uncertainty (accuracy) ±1 dB
Frequency range (17 MHz to 30 MHz)
Frequency resolution 1 kHz steps
Frequency uncertainty (accuracy) ±(50 ppm + 100 Hz)
Level range –20 dBm to 0 dBm at 100 1
Level resolution 0.1 dB
Level uncertainty (accuracy) ±1 dB
Impedance 100 1, 120 1, 135 1, 150 1, 600 1
Receiver characteristics Reception frequency range 200 Hz to 20 kHz
20 kHz to 30 MHz
Frequency uncertainty (accuracy) ±(50 ppm + 1 digit)
VF reception level range –90 dBm to 15 dBm at 600 1
VF level uncertainty (accuracy) 200 Hz to 20 kHz
–90 dBm to –50 dBm, uncertainty (accuracy) ±2 dB
–50 dBm to 15 dBm, uncertainty (accuracy) ±1 dB
WB reception level range –80 dBm to 15 dBm at 100 1, 120 1,135 1, 150 1
WB level uncertainty (accuracy) 20 kHz to 2.2 MHz
–80 dBm to –50 dBm, uncertainty (accuracy) ±2 dB
–50 dBm to 15 dBm, uncertainty (accuracy) ±1 dB
2.2 MHz to 30 MHz
–80 dBm to –50 dBm, uncertainty (accuracy) ±2 dB
–50 dBm to 15 dBm, uncertainty (accuracy) ±1 dB
Impedance 100 1, 120 1, 135 1, 150 1, 600 1
POTS dialer DTMF 0 – 9, #, *
Phonebook 25 entries
Digital multimeter (DMM) Test type Snapshot and continuous
Impedance selection (for voltage measurement) 100 k1, 1 M1, 10 M1