Exfo FTB-610 Wideband Copper Test Module Manual de usuario Pagina 343

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 352
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 342
Pruebas de cobre 335
A Especificaciones técnicas
IMPORTANTE
Las siguientes especificaciones técnicas pueden cambiar sin previo
aviso. La información contenida en esta sección se proporciona
únicamente como referencia. Si desea obtener las especificaciones
técnicas más recientes del producto, visite la página web de EXFO
en www.exfo.com
COPPER SPECIFICATIONS
a, b, c
Transmitter characteristics
Frequency range (200 Hz to 20 kHz)
Frequency resolution 1 Hz steps
Frequency uncertainty (accuracy) ±(50 ppm + 1 Hz)
Level range –20 dBm to 0 dBm at 600 1
Level resolution 0.1 dB
Level uncertainty (accuracy) ±1 dB
Frequency range (20 kHz to 2.2 MHz)
Frequency resolution 1 kHz steps
Frequency uncertainty (accuracy) ±(50 ppm + 100 Hz)
Level range –20 dBm to 0 dBm at 100 1
Level resolution 0.1 dB
Level uncertainty (accuracy) ±1 dB
Frequency range (2.2 MHz to 17 MHz)
Frequency resolution 1 kHz steps
Frequency uncertainty (accuracy) ±(50 ppm + 100 Hz)
Level range –20 dBm to 0 dBm at 100 1
Level resolution 0.1 dB
Level uncertainty (accuracy) ±1 dB
Frequency range (17 MHz to 30 MHz)
Frequency resolution 1 kHz steps
Frequency uncertainty (accuracy) ±(50 ppm + 100 Hz)
Level range –20 dBm to 0 dBm at 100 1
Level resolution 0.1 dB
Level uncertainty (accuracy) ±1 dB
Impedance 100 1, 120 1, 135 1, 150 1, 600 1
Receiver characteristics Reception frequency range 200 Hz to 20 kHz
20 kHz to 30 MHz
Frequency uncertainty (accuracy) ±(50 ppm + 1 digit)
VF reception level range –90 dBm to 15 dBm at 600 1
VF level uncertainty (accuracy) 200 Hz to 20 kHz
–90 dBm to –50 dBm, uncertainty (accuracy) ±2 dB
–50 dBm to 15 dBm, uncertainty (accuracy) ±1 dB
WB reception level range –80 dBm to 15 dBm at 100 1, 120 1,135 1, 150 1
WB level uncertainty (accuracy) 20 kHz to 2.2 MHz
–80 dBm to –50 dBm, uncertainty (accuracy) ±2 dB
–50 dBm to 15 dBm, uncertainty (accuracy) ±1 dB
2.2 MHz to 30 MHz
–80 dBm to –50 dBm, uncertainty (accuracy) ±2 dB
–50 dBm to 15 dBm, uncertainty (accuracy) ±1 dB
Impedance 100 1, 120 1, 135 1, 150 1, 600 1
POTS dialer DTMF 0 – 9, #, *
Phonebook 25 entries
Digital multimeter (DMM) Test type Snapshot and continuous
Impedance selection (for voltage measurement) 100 k1, 1 M1, 10 M1
Vista de pagina 342
1 2 ... 338 339 340 341 342 343 344 345 346 347 348 ... 351 352