Working with iOLM files
Post Processing Software 247
Customizing the Elements Table Tab
To change the launch fiber and receive fiber positions:
1. Open the iOLM file.
2. Enter the launch fiber length and receive fiber length in the Test
Settings table under the Summary tab.
When the launch fiber length and receive fiber length are changed, a
partial analysis is performed. In this partial analysis, the values for link
length, link loss, link ORL, element position, element section length, and
diagnostics are recalculated. This partial analysis does not change the
elements you have added or the elements you have changed in the link.
Note: If the entered launch fiber length and receive fiber length are not valid, the
application prompts an error message. If no element is present at the
position, then a new element is automatically inserted in that place.
Note: If a macrobend is set either as link start or link end, a new connector is
added and that element will be set as link start or link end.