Exfo PSO-200 Optical Modulation Analyzer Manual de usuario Pagina 153

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 285
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 152
SCPI Commands Reference
Optical Modulation Analyzer 145
Product-Specific Commands—Description
:SENSe:DDEMod:AVERage:TCONtrol
Description This command sets the average mode that is to
be managed, when the number of scans
processed exceeds the Average Number of
bursts.
At *RST, this value is LAS.
Syntax :SENSe:DDEMod:AVERage:TCONtrol<wsp>EXP
onential|LAST
Parameter(s) Avg:
The program data syntax for the first parameter is
defined as a <CHARACTER PROGRAM DATA>
element. The allowed <CHARACTER PROGRAM
DATA> elements for this parameter are:
EXPonential|LAST.
Selects the average mode to be set.
EXPonential, new results are averaged
exponentially.
LAST, last N bursts will be used to compute the
average.
Example(s) SENSe:DDEMod:AVERage:TCONtrol LAS