Exfo PSO-100 Series Optical Sampling Oscilloscopes Manual de usuario Pagina 122

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 140
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 121
Measurement Definitions
114 PSO-100
Eye Height
Eye Height
The eye height (EH) is a measurement of the vertical opening of an eye
diagram and it is defined as:
Eye Amplitude
The eye amplitude (EA) is simply defined as the difference between the
one and zero levels according to:
Eye Opening Factor
The eye-opening factor is a measurement of the vertical eye opening
according to:
EH P
one
3σ
one
()P
zero
3σ
zero
+()=
EA P
one
P
zero
=
EOF
P
one
σ
one
()P
zero
σ
zero
+()
P
one
P
zero
----------------------------------------------------------------------------
=
Vista de pagina 121
1 2 ... 117 118 119 120 121 122 123 124 125 126 127 ... 139 140

Comentarios a estos manuales

Sin comentarios