Exfo PM-1100 Optical Power Meter Manual de usuario Pagina 3

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 4
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 2
P/N: 1051923 January 2008 3/4
³
The SENS:FREQ:CAT? query should be named and defined as follows:
³ The SENS:POW:WAV command should be named and defined as
follows:
³ The SENS:POW:WAV? query should be named and defined as follows:
SENSitivity:FREQuency:CATAlog?
Description This query returns a list of available sampling rates.
Syntax SENS:FREQ:CATA?
Response List of available sampling rates in the
“40.0;20.0;10.0;5.0;1.0;0.5;0.1” format.
Example SENS:FREQ:CATA?
SENSitivity:POWer:WAVElength
Description This command selects a new operating wavelength.
Syntax SENS:POW:WAVE<space><numeric_value> [<space>NM]
Parameters The <numeric_value> is an operating wavelength expressed in
nanometers (nm). Any wavelength within the spectral range of
the optical detector (at 1 nm resolution) may be selected. See
the section on optical specifications in the user guide for the
exact spectral range of each detector type.
Example SENS:POW:WAVE 1310
SENSitivity:POWer:WAVElength?
Description This query returns the currently selected calibrated wavelength.
Syntax SENS:POW:WAVE?
Response The current wavelength in nanometers (nm) in the “9999 nm”
format.
Example SENS:POW:WAVE?
Vista de pagina 2
1 2 3 4

Comentarios a estos manuales

Sin comentarios